An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Photoelectron Spectroscopy • Lecture 5 – instrumental details • General spectrometer design • Vacuum generation and measurement

  2. Electron Amplifier, Counter, and Recorder Photon Source Electron Detector Electron Kinetic Energy Analyzer Sample Vacuum Pumps Required spectrometer components

  3. Spectrometer design considerations • What type of samples are of interest? • What resolution is required to gather the information of interest? • Resolution of this experiment is always instrumental based. • Experimental resolution is directly correlated with electron kinetic energy. • What can be done to increase experimental sensitivity? • Often a pay-off between resolution and sensitivity. • What is it going to cost? • Money for constructing spectrometer • Pumping requirements • Time for data collection

  4. Sample Considerations • Gas-Phase Photoelectron Spectroscopy • Atoms, neutral Molecules, anions, clusters, etc. • For neutral molecules, need a vapor pressure of ≈10-4 torr in high vacuum at temperatures <≈ 500 °C • “Hot” molecules will have more complicated vibrational/rotational contributions. • Condensed-Phase Photoelectron Spectroscopy • Film on conductive surface • For valence spectroscopy, need uniform film • (vapor deposition, SAMs, spin coating)

  5. Why do we need vacuum? • Low pressure is required for operation of electron detectors • Pressure must be low enough to allow mean-free-path of electrons through the analyzer • Pressure must be low enough that gas-phase samples are volatile • Ultra-high vacuum is required to lower surface contamination for condensed-phase spectroscopy • Vacuum pump: a pump that removes gas molecules from a sealed volume in order to leave behind a partial vacuum

  6. Vacuum Ranges

  7. Examples of vacuum levels

  8. How Low Must Pressure be for a Surface to be “Clean”? If “sticking coefficient” S = 1 And pressure = 2.5 x 10-6 Torr A monolayer will form in 1 second Lower pressure to ~ 10-9 Torr A monolayer forms in 1,000 seconds S is usually <<1

  9. Methods for Vacuum Generation: 1 • Positive displacement: use a mechanism to repeatedly expand a cavity, allow gases to flow in from the chamber, seal off the cavity, and exhaust it to the atmosphere (rotary vane, scroll pump, roots blower)

  10. Methods for Vacuum Generation: 2 • Momentum transfer: use high speed jets of fluid or rotating blades to knock gaseous molecules out of the chamber (diffusion, turbomolecular)

  11. Methods for Vacuum Generation: 3 • Entrapment: capture gases in a solid or absorbed state (cryopumps, getters, ion pumps)

  12. Vacuum measurement Bourdon gauge > 10-2 torr McLeod gauge > 10-4 torr thermocouple gauge 760 - 10-3 torr ion gauge 10-3 - 10-10 torr