An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Qb(214Bi)=3.2 Me Bi-Po Process 238U 214Po (164 ms) b 214Bi (19.9 mn) a 210Pb 22.3 y Scintillator + PMT 0.021% 210Tl (1.3 mn) Tracking (wire chamber) 232Th Source foil (40 mg/cm2) 212Po (300 ns) b 212Bi (60.5 mn) Shield radon, neutron,g a 208Pb (stable) 36% 2 modules 23 m2→ 12 m2 Background < 1 event / month e- a delay 208Tl (3.1 mn) R&D BiPo DETECTOR To measure the purity in 208Tl and 214Bi of the bb source foils before the installation in SuperNEMO Goal: To measure 5 kg of foils (12 m2, 40 mg/cm2) in 1 month with a sensitivity of: 208Tl < 2 mBq/kg and 214Bi < 10 mBq/kg

  2. e- prompt e- Bi-Po Process 238U 214Po Qb (212Bi) = 2.2 MeV a (164 ms) b 214Bi (19.9 mn) Delay a a 210Pb 22.3 y T1/2 ~ 300 ns Edeposited ~ 1 MeV Scintillator + PMT 0.021% 210Tl (1.3 mn) Tracking (wire chamber) 232Th Source foil (40 mg/cm2) 212Po (300 ns) b 212Bi (60.5 mn) Shield radon, neutron,g a 208Pb (stable) 36% 2 modules 23 m2→ 12 m2 Background < 1 event / month 208Tl (3.1 mn) R&D Detecteur BiPo To measure the purity in 208Tl and 214Bi of the bb source foils before the installation in SuperNEMO Goal: To measure 5 kg of foils (12 m2, 40 mg/cm2) in 1 month with a sensitivity of: 208Tl < 2 mBq/kg and 214Bi < 10 mBq/kg

  3. With 5 kg of 82Se source foil (~ 12 m2, 40 mg/cm2) 2 mBq/kg of 208Tl 50 (e-, delay a) 212Bi decays / month 3 events / month e ~ 6% Possible design Two modules, each module 2 x 3 m2 Calorimeter: 2 x 150 PMTs + Scint. Blocks 2 x 20 x 20 cm2 15 scint. Bars 2 m long, 2 x 20 cm2 Tracking: 4 layers of Geiger drift cells Magnetic field suitable to reject external e- Room needed: 4 x 5 m2 (with shield) + clean room beside the detector

  4. quenching a energy (MeV) Fit between 40 and 130 ns : T 1/2 = (212 +/- 65) ns ~ 300 ns expected Qb ~ 2.2 MeV Time delay between a and electron (in ns) electron energy (MeV) Analysis of such events in NEMO-3 data 1642 events obvserved in 1 year of data If all comes from mylar: 2.5 mBq/kg a 40 ns < Tdelay < 130 ns e- T0 electron (trigger)

  5. Origin of backgrounds surface contamination of 208Tl on the entrance surface of the lower scintillator Two prototypes to study the level of background

  6. Prototype BiPo-1 end 2005 – Jun. 2006 Air outlet + cables Surface of scintillators: Spottering of very thin layer of metal on the surface of the scintillators: 100 nm internal Al for reflecting + 100 nm external Au for cleaning PMT 5" NEMO3 Light guide « screan » to stop scintillation light Naked scintillator 20x 20 x 2cm We will use NEMO-3 equipments (5” PMTs, scintillator, etc…) 1 cm bored Polyethylen lead 10 cm bored Polyethylen Radon-tight enveloppe Radon-free air

  7. Prototype BiPo-2 Proto 2: Jun. 2006 – Jun. 2007 1 x 1 m2→ 25 x 2 PMTs (20 x 20 cm2)

  8. BiPo detector may become a new low background detector (like HPGe generation…) to measure 214Bi and 208Tl purity of thin materials (volume or surface) e- bb foils for SuperNEMO Capton foils for GERDA Cables ? etc… a Limitation of the thickness: 212Bi electron must cross the material • Final design must be studied depending on what we want to measure • The detector could be installed in Canfranc ?…