1 / 19

A UNIVERSAL PROGRAMMABLE TEMPERATURE SPRAY CHAMBER

A UNIVERSAL PROGRAMMABLE TEMPERATURE SPRAY CHAMBER. Jerry Dulude, Glass Expansion, jdulude@geicp.com . Limitations of Room Temperature Spray Chambers. Intensity drift with temperature Excessive plasma loading (volatile solvents) Excessive oxide formation

shoushan
Download Presentation

A UNIVERSAL PROGRAMMABLE TEMPERATURE SPRAY CHAMBER

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. A UNIVERSAL PROGRAMMABLE TEMPERATURE SPRAY CHAMBER Jerry Dulude, Glass Expansion, jdulude@geicp.com

  2. Limitations of Room Temperature Spray Chambers • Intensity drift with temperature • Excessive plasma loading (volatile solvents) • Excessive oxide formation • Insufficient control of analyte transport Glass Expansion www.geicp.com 1

  3. Limitations of Externally Controlled Spray Chambers • Requires antifreeze solution • Requires bulky floor model chiller • Messy coolant lines from chiller to chamber • Freezing of condensate on lines • Requires intricate jacketed chamber Glass Expansion www.geicp.com 1

  4. IsoMist™ Controlled Temperature Spray Chamber Glass Expansion www.geicp.com 1

  5. IsoMist Encapsulated Spray Chamber Glass Expansion www.geicp.com 1

  6. Features of IsoMist • Powerful Peltier-effect chiller reaches -5C in 15 minutes • Programmable from -10 to 60C in 1 degree increments • Maintains temperature to within 0.1 degree • Custom software runs from any PC • Compact design (7.5x4x4 inches) • 100% self-contained (no external lines) • Incorporates Bluetooth® technology for clean wireless control (USB available) Glass Expansion www.geicp.com 1

  7. PC Screen showing IsoMist Software Glass Expansion www.geicp.com 1

  8. Applications of the IsoMist • Analyze volatile solvents • Reduce oxide interferences • Lower detection limits • Increase stability Glass Expansion www.geicp.com 1

  9. Determination of Trace Metals in Naphtha • Interfere with the cracking process • Poisoning of the catalysts • Environmental release concerns • Origin and migration markers Glass Expansion www.geicp.com 1

  10. Analysis of Naphtha by ICP-OESPE Optima 2100 DV (Axial view) • Forward Power: 1500Watts • Coolant flow: 20L/min • Auxiliary flow: 1.8L/min • Nebulizer gas flow: 0.35L/min • Injector: 1mm capillary bore • Spray Chamber: Twister baffled cyclonic • Nebulizer: SeaSpray glass concentric • Uptake rate: 300ul/min • IsoMist temperature: -10C Glass Expansion www.geicp.com 1

  11. IsoMist on Optima 2100DV Glass Expansion www.geicp.com 1

  12. Spike Recoveries in Naphtha Glass Expansion www.geicp.com 1

  13. Oxide Reduction in ICP-MSIsobaric oxide interferences Robert Thomas, Practical Guide to ICP-MS, Marcel Dekker, NY, 2004. Glass Expansion www.geicp.com 1

  14. Reducing Oxides for ICP-MS Data courtesy of David Jones, ALS Chemex, Brisbane, Australia Glass Expansion www.geicp.com 1

  15. Analysis of Limited Sample Volumes • Biological applications • Intracellular fluids • Neonatal samples • Forensic applications • Requires micro flow nebulization • Low DL’s desirable Glass Expansion www.geicp.com 1

  16. Effect of Temperature on Normalized Intensity (20ul/min Uptake) ICP-OES PE Optima 2100DV Glass Expansion www.geicp.com 1

  17. Effect of Temperature on DL20ul/min Uptake (ICP-OES) PE Optima 2100DV Glass Expansion www.geicp.com 1

  18. Without IsoMist IsoMist At 21C Stability EnhancementICP-OES Glass Expansion www.geicp.com 1

  19. PE ICP & ICP-MS Varian ICP & ICP-MS Agilent ICP-MS ThermoFisher ICP & ICP-MS Horiba J-Y ICP Teledyne Leeman ICP GBC ICP Spectro ICP Borosilicate glass Quartz Inert PFA IsoMist Compatibility Models Spray Chamber Materials Glass Expansion www.geicp.com 1