1 / 7

ABG Instrumentation Analytical Errors

ABG Instrumentation Analytical Errors. Levy-Jennings Chart In Control. +2SD. X. -2SD. Random Error. +2SD. X. -2SD. Contributing Factors Statistical probability Contamination of the sample Improper handling of the sample. Corrective Actions

amil
Download Presentation

ABG Instrumentation Analytical Errors

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. ABG InstrumentationAnalytical Errors

  2. Levy-Jennings ChartIn Control +2SD X -2SD

  3. Random Error +2SD X -2SD

  4. Contributing Factors Statistical probability Contamination of the sample Improper handling of the sample Corrective Actions Reanalyze control sample after performing a 2-point calibration If this works no further action is needed If this does not work, follow manufacturer recommendations Random Errors

  5. Systematic or Bias ErrorTrend +2SD X -2SD

  6. Systematic or Bias ErrorShift +2SD X -2SD

  7. Contributing Factors Contaminated buffer solutions Incorrect gas concentrations Component degradation/ failure Incorrect procedure Corrective Actions Troubleshoot suspected problems Repair or replace problematic components Re-run 2 point calibration Bias errors