Capillary Electrophoresis

Capillary Electrophoresis. Separation of analyte ions via differential migration in an electric field, coupled with electro-osmotic flow of mobile phase. Advantages Only needs nL sample High speed and resolution, virtually no band broadening Instrumentation

celine
Download Presentation

Capillary Electrophoresis

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Capillary Electrophoresis Separation of analyte ions via differential migration in an electric field, coupled with electro-osmotic flow of mobile phase Advantages • Only needs nL sample • High speed and resolution, virtually no band broadening Instrumentation • Capillary tube (10 - 100mM internal diam., 40-100cm long) • Two buffer reservoirs, with platinum electrodes • DC potential (20-30 kV) applied along capillary • Sample introduced one end, detector at other • Direct of potential depends on charge (+/-) of analyte

  2. Capillary Electrophoresis Mobile Phase • Commonly phosphate or borate buffer (20-100 mM) • pH and Ionic strength must be controlled • Can add detergents to transport neutral molecules in a micelle (MEKC) Stationary Phase • No stationary phase for true CE • Newer developments introducing a stationary phase combine CE and HPLC to give electrochromatography Principles of separation Based on interaction of analyte with electric field Migration velocity v = (µe + µeo) E where µeand µeoare the electrophoretic mobilities of the analyte and buffer, and E is the applied field strength

  3. Capillary Electrophoresis Retention and Resolution: dependent on • Charge / size ratio is primary separation factor • Charge gives v  and thus RT  • Size gives v  and thus RT  • Interaction with buffer ions / molecules • pH and ionic strength of buffer affects ionisation of analyte – and thus RT • Applied field • Column Length • Diffusion Optimisation: may involve • bufferpH - alter charge status of analyte • buffer ionic strength - change capacity to produce electro-osmotic flow • organic modifiers: influence ionic strength, and can “capture” analytes selectively to affect charge/size ratio

  4. Capillary Electrophoresis Detection: normally similar to LC detectors but • Peak area is dependent on rate of movement through the detector • Peak area not independent of retention time • Mostly “on-column”, i.e. capillary forms flow cell • Short path length gives reduced sensitivity • Indirect methods may be needed to increase sensitivity Applications There are a variety of named techniques, each suitable for different analytical problems. You may see mentioned: • Capillary zone electrophoresis • Capillary gel electrophoresis • Capillary isotachophoresis • Capillary isoelectric focusing • Capillary electrochromatography • Micellar electrokinetic capillary chromatography