1 / 11

Data Acquisition System for NPDGamma

Data Acquisition System for NPDGamma. Nadia Fomin April 4 th , 2008 Bon Appetit!. HAZEL/DAQ. VME 1. VME 2. VME 3. You. DAQ!. CLOVER/CONTROL. SSH. SSH. SSH. SSH. 4 rings – 12 Cesium Iodide detectors each. (1 ≤ i ≤ 12) (1 ≤ j ≤ 4). Detector Signal - S j i.

chipo
Download Presentation

Data Acquisition System for NPDGamma

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Data Acquisition System for NPDGamma Nadia Fomin April 4th, 2008 Bon Appetit!

  2. HAZEL/DAQ VME 1 VME 2 VME 3 You DAQ! CLOVER/CONTROL SSH SSH SSH SSH

  3. 4 rings – 12 Cesium Iodide detectors each (1 ≤ i ≤ 12) (1 ≤ j ≤ 4) Detector Signal - Sji • Instead of storing the signal from each detector for every neutron pulse: • <Sj>=> Average signal for a ring =4 SUM Signals=>VME3 • Sdiff,ij => Sij -<Sj>=> 48 DIFFERENCE Signals => VME2 • 100 time bins for SUM and DIFFERENCE signals, with each time bin summed over 20 (DIFFs) and 25 (SUMS) samples

  4. 4 rings – 12 Cesium Iodide detectors each (1 ≤ i ≤ 12) (1 ≤ j ≤ 4) Detector Signal - Sji • Instead of storing the signal from each detector for every neutron pulse: • <Sj>=> Average signal for a ring =4 SUM Signals=>VME3 • Sdiff,ij => Sij -<Sj>=> 48 DIFFERENCE Signals => VME2 • 100 time bins for SUM and DIFFERENCE signals, with each time bin summed over 20 (DIFFs) and 25 (SUMS) samples

  5. VME 1 VME 2 VME 3 Proton Current Difference Signals Sum Signals RFSF Voltage and Current • => Data from each pulse is written to 3 files, • corresponding to the VME that it came from. • => Data from each pulse is preceded by a header • (in each of the 3 files), including: • Date and time • Time (us) since last pulse (better be 50ms) • Pulse number • DIO data -> most meaningful in VME3, which records the spin sequence • Time since last data write and number of records written Monitor Data (3)

  6. TO 1.1 2.1 3.1 4.1 Fiberoptic Link Receiver TTL N FIFO FIFO TTL NIM Gate/Delay Generator a b delay 1.2 3.2 TTL C TTL Gate/Delay Generator NIM 1.3 3.3 b 1us b TTL N Gate/Delay Generator NIM ?a a gate 1.4 3.4 TTL C gate Gate/Delay Generator NIM Pins 4 Scanclock gate VME 1.3 DIO VME 1.4 Alphi VME 1.? Joerger VME 2 Pins 2 Clock out

  7. T0 Copy to VME2 trigger reset Gate ff 1.1 2.1 3.1 5.1 A B 6.1 4 TTL NIM RESET Gate/Delay Generator 4-fold Logic Gate ttl T0 FIFO Gate/Delay Generator trigger a 1.2 5.2 6.2 b N Gate/Delay Generator b TTL 4-fold Logic Gate NIM reset c TTL c out delay 1.3 2.3 Gate 80 ns 5.3 3.2 b TTL C FIFO A Gate/Delay Generator NIM b B Gate/Delay Generator nim a a 1.4 5.4 TTL TTL N Gate/Delay Generator NIM c Spin Seq Out d c SF out gate d gate Note: Wire on right VME 1.3 DIO VME 1.4 Alphi VME 1.7 Joerger WF GEN VME 1.8 Joerger RFSF T0 8 14 | 17 Scanclock 6 4 T Clock out 2 12 VME 3

  8. b a 1.1 2.1 3.1 4.1 5.1 6.1 10.1 8.1 N TTL NIM COINC FIFO FIFO COINC Gate/Delay Generator b b Gate/Delay Generator Gate/Delay Generator 1.2 2.2 3.2 5.2 6.2 7.2 C TTL FIFO FIFO COINC COINC Gate/Delay Generator NIM b TTL A 8 us a 1.3 a 3.3 5.3 6.3 8.2 4.2 C TTL NIM FIFO FIFO Gate/Delay Generator RFSF T0 Gate/Delay Generator Gate/Delay Generator Gate – 1us a 1.4 a 3.4 5.4 6.4 In - ? N TTL NIM FIFO FIFO Gate/Delay Generator b Gated T0 Out – 10 ns wide b ttl 9.1 1 us - wide Ext. Trig Gate/Delay Generator Digital Delay Pulse Generator a 60Hz Ref b nim Pins 14 9.2 COINC Joerger Scalar VME 1.3 DIO VME 1.5 LeCroy 1182 VME 1.6 LeCroy 1182 10.2 Gate/Delay Generator Arm In T0 1 1 us delay TTL 2 3 Pins 2 VME 1

  9. Current Status: • After some changes, VME1 and VME3 take data (huge improvement from last week when none worked) • VME2 is not completely operational (takes data even when T0 cable is unplugged) • Once it’s up and running, a noise measurement will be done, and then work on the 60hz “upgrade” will seriously begin SNS – 60 Hz operation Doing the naïve software switch, and scaling 20Hz->60Hz, VME2 (largest volume of data) can’t keep up. Issue -> DMA transfer, ADCs can’t read/write simultaneously

  10. Ideas : • Figure out the minimum amt of time that VME2 needs to read out the ADCs, see if that can be found during the SNS pulse. • Problem: All neutrons are slow and probably useful. Upside: If enough time, no money or significant coding/procedural changes. • Add more memory to the ADCs, so that 8 pulses can be stored (one octet), read out every 9th pulse. • Problem: Costs money. Upside: Lose few good pulses. • BUT! SNS pulses are shorter, so 8 might already fit! • 3. Possibly alter the spin sequence, so that every nth (3rd?) sequence is garbage and use that time for data transfer. So, we'd still have an octet, but it would be a subset of 12 spin sequences. Problem: Throw away a third of good pulses. • Example: (1,0,0,1,0,1,1,0)-> (1,0,X,0,1,X,0,1,X,1,0,X) • 4. Installing a real-time kernel. Not sure how that'll fix the problem.