300 likes | 545 Views
BMC Medical Co., Ltd. Training Course. Nov. 2011. Table of Contents CATALOGUE INSTALL& DISASSEMBLE SET UP. 1. CATAOLGUE. Resmart CPAP. 2. INSTALL & DISASSEMBLE. Install Water tank. Component : Water tank, Humidifier, Device. Clearing Water tank.
E N D
BMC Medical Co., Ltd. Training Course Nov. 2011
Table of Contents CATALOGUE INSTALL& DISASSEMBLE SET UP
Clearing Water tank • Disconnect the tubing from the water chamber. Press down on the water chamber and slide it out of the humidifier platform. • Open the silica gel cover of water chamber and discard any remaining water. • Fill a solution of warm water and a mild dishwashing detergent into the chamber, cover the silica gel cover, rock the chamber a few minutes, then outflow the solution. Rinse the chamber several times with clean water and allow to air dry. • Fill the water chamber and close the silica gel cover. Inspect the water chamber for any leaks or damage. • Replace the water chamber if any damage is present.
DISASSEMBLE • Open the cover of the filter carefully. And then take it out.
DISASSEMBLE • Press the button at the bottom of the main machine, then pull the heated humidifier.
SET UP • Control Button
CPAP • CPAP USER SET UP ENTER USER MENU
CPAP SET UP • Verify/Change the RESmart USER settings. To start thesetup menu, press and hold the Ramp button until the setup menu appear (about 5 seconds) when system standby. • IMPORTANT! Press the Ramp Button to enter change status and press +/- User Buttons to change the setting. Press Ramp Button again to confirm the setting. • Press +/- User Buttons to access the next setting.
CPAP • Setting the Humidifier When the optional InH2™ Heated Humidifier is used, this setting can change the temperature and humidity of airflow between level 0 to 5. Default setting is 3.
CPAP • Setting the Altitude • The altitude setting can be changed between level 0 to 2. Default setting is 0. • 0 = less than 2,460 ft. (<750 m) • 1 = 2,460 to 4,921 ft. (750 m to 1500 m) • 2 = 4,924 to 8,202 ft. (1501 m to 2500 m) • *over 8,202 ft. = The airflow pressure may not be accurate. Contact your home care provider to have your pressure setting adjusted.
CPAP • Setting Auto ON • When the RESmart CPAP is standby and wearing mask, your deep breath will start airflow automatically. Default setting is Enable.
CPAP • Setting Auto Off • When the mask is removed, the RESmart CPAP will stop airflow automatically. Default setting is Disable. • Note:We advice you can set Auto Off Disable. Most machine may appear this kind of problem, because apnea and mask off is not to be judged at pressure lower than 7 hpa
CPAP • Setting the LCD Light • This setting allows you to adjust the LCD light among Auto, On and Off. Default setting is Auto. • Auto: LCD light will be turned on when any button pressing and turned off after few seconds. • On: LCD light will be turned ON always. • Off: LCD light will be turned OFF always.
CPAP • Setting the Alert • This setting allows you to turn on/off the patient disconnect alert. When a large, continuous air leak (such as mask removal) has been detected in the circuit, this setting enables/disables the audible alert (a beeping sound). This setting is only effective when the Auto Off setting is disabled. (This function is NOT available on certain model.) Default setting is Disable.
CPAP • Setting the Delay Off • When the optional InH2™ Heated Humidifier is used, this setting allows the air flow continue for about 15 minutes at very low pressure (about 2hPa) after RESmart CPAP is turned off. This will blow off the left vapor and protect the RESmart CPAP. Default setting is Enable.
CPAP • Use Days • The number of nights the RESmart CPAP was used for therapy for more than 4 hours will appear. This screen is only for reference. Your home care provider may periodically ask you for this information.
CPAP • Exit the Settings • The settings are complete. Press the +/- user button to access the SAVE setting and Ramp Button to save all changes and exit the settings menu.
CPAP • CPAP MAINTENANCE SET UP ENTER MAINTENANCE MENU
CPAP • Entering the Menu: Press and hold the Pressure Start/Stop Button and Ramp Button while connecting the power cord into the AC inlet. After several seconds the Maintenance Menu will appear on display screen as below. • IMPORTANT! Press the Ramp Button to enter change status and press +/- User Buttons to change the setting. Press Ramp Button again to confirm the setting. • Press +/- User Buttons to access the next setting as below.
CPAP • Ramp Starting Pressure: The starting pressure after blower is turned on. Setting range is 4hPa to Treatment Pressure. Adjustment step is 0.5cm H2O. Default setting is 4cm H2O.
CPAP • Treatment Pressure: The prescription pressure after ramp time. Setting range is 4-20cm H2O. Adjustment step is 0.5cm H2O. Default setting is 8cm H2O.
CPAP • Ramp Timer: The ramp time from starting pressure to prescription pressure. Setting range is 0-60mins. Adjustment step is 5mins. Default setting is 15mins.
CPAP • Manual Pressure: This setting allows user to adjust Treatment Pressure by oneself. Setting range is 0-2cm H2O. Adjustment step is 0.5cm H2O. Default setting is 0. • For example: if this setting is 1cm H2O and Treatment Pressure setting is 8cm H2O, user may change the treatment pressure between 7-9cm H2O by pressing +/- User Button.
CPAP • Use Time: Total hours this RESmart has been used will appear. This number is only for reference.
CPAP • Save and Exit: The settings are complete. Press the +/- user button to access the SAVE setting and Ramp Button to save all changes and exit the Maintenance Menu. • IMPORTANT! To make all changes active, press the +/- User Button to access the SAVE setting and Ramp Button to save all changes and exit the Maintenance Menu.