An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Results from the CUORE experiment Antonio Branca– DFA UniversitàdegliStudi di Padova & INFN Sezione di Padova On behalf of the CUORE Collaboration EPS-HEP 2019 – July 10-17, Ghent, Belgium

  2. Outline • Bolometric technique for neutrino-less double beta decay search; • Expected signature from ; • Why TeO2? • CUORE: the first bolometric experiment at the ton scale: • The challenge of a rare event search; • Experimental setup; • From detector construction to data taking: • Assembly/installation; • Cooldown and commissioning ; • Detector optimization and performance; • Data-taking; • Physics results: • Neutrino-less DBD analysis; • Background budget; • Two neutrino DBD analysis; • Summary A. Branca

  3. The expected signature • Building a high sensitivity experiment: • : select candidates with high natural isotopic abundance or enriched; • : high detection efficiency; • : high detector mass; • : good detector stability over a long period; • : extremely high energy resolution; • : extremely low background environment; Ideal energy spectrum of the two electrons in the final state A. Branca

  4. Bolometric technique A (A) Copper frame:10 mK heat sink (B) PTFE holders:weak thermal coupling D Radiation: energy deposit C (D) Si joule heater:reference pulses B E (C) TeO2 crystal:energy absorber (E) NTD Ge thermistor: resistive thermometer Low temperature needed: Readout A. Branca

  5. Bolometric technique: 130Te • Bolometric detectors: detector also the source of : • high efficiency; • excellent energy resolution; • large masses are achievable; • 130Te: a good candidate source for : • high natural isotopic abundance; • Q-value (2528 keV) above most of the natural radioactivity; • nuclear matrix elements and phase space on average; A. Branca

  6. A rare event search Searching for a rare event (): ……………………… Extremely important to reduce as much as possible backgrounds: • natural radioactivity from outside the detector: • cosmic ray muons induced background; • neutron and gamma fluxes; • natural radioactivity from the detector itself: • long-lived nuclei (40K, 238U, 232Th); • anthropogenic radioactive isotopes (60Co, 137Cs, 134Cs); • cosmogenic radioactive isotopes (60Co); • mechanical vibration noise: • cryogenic system and seismic noise; CUORE installed @ LNGS underground laboratories (~3600 m.w.e.) REDUCTION • Strict protocols for crystal production @ SICCAS; • Cleaning techniques developed @ LNL for copper parts near crystals; • Strict protocol for assembling and installation; REDUCTION Suspension/damping systems and new noise cancellation tools REDUCTION A. Branca

  7. Suspension system and external shielding Y-Beam • Abatement of vibrations: mechanical decoupling from the outside environment: • detector suspension independent from that of cryogenic and calibration systems: • detector hung by the Y-Beam through cables made of stainless steel tie bars, Kevlar ropes and copper bars (damping the horizontal oscillations); • 3 minus-K springs connect the Y-Beam to the Main Support Plate, MSP (attenuating the noise of ~35 dB); • elastomers at the structure basis (seismic isolators); • Reduction of radioactive background (from outside): • outer neutron shield: polyethylene + boric acid; • outer gamma shield: lead shield; Minus-K MSP External lead shield (~70 t) H3BO3 panels Polyethylene Elastomers A. Branca

  8. Cryogenic system and shielding • Challenging task: cool down ~15 tons @ T < 4 K and ~1.5 tons @ T = 10 mK in a few weeks in a low radioactive environment. • Cryogen-free (dry) cryostat: high duty cycle: • Fast Cooling System (4He gas): T down to ~40 K; • 5 Pulse Tubes cryocoolers: T down to ~4 K; • (Custom) Dilution Refrigerator: T operations 10 mK; • Nominal cooling power: 3 μW @ 10 mK; • Reduction of radioactive background (from detector): • material screening and accurate selection to ensure radiopurity (mainly pure copper, other material in small amount, limited amount of Multi Layer Insulator); • cleaning of copper surfaces facing crystals; • lead shielding (Roman and modern Pb); A. Branca

  9. Cryostat commissioning and tower assembly • Cryostat commissioning completed in March 2016: • stable base T = 6.3 mK over 70 days (no detector, full load); • full detector read-out chain (electronics, DAQ) test, temperature stability with Mini-Tower (8 crystal tower); • successful deployment of the calibration sources at base temperature; • Strict protocol adopted for each step of towers assembling/installation (developed thanks to CUORE-0): • assembling:in N2 atmosphere and within glove boxes to avoid radioactive recontamination; • installation: protected area inside clean room flushed with radon free air; protective bags flushed with N2 for overnight and emergency storage; A. Branca

  10. Detector installation • Design specifics: • detector arrangement: 19 towers with 13 floors of 4 crystals each; • crystals: 988 crystals, 5 cm3, 750 g each; • total TeO2 mass of 742 kg; • total 130Te mass of 206 kg (all natural abundance); • Reduction/control of radioactive background: • minimization of material/surface facing the crystals; • closely packed crystal array with high granularity; • GOALS: • low background of ; • energy resolution: in the Region Of Interest (ROI); • projected sensitivity: (5 years, 90% C.L.); Installation of all 19 towers completed by August 2016 A. Branca

  11. Cooldown and temperature selection • Temperature scans around base temperature to optimize detector resolution and NTDs resistances at design values (): • March 2017: identified best working temperature of 15 mK; • September 2017: with calibration sources deployed, confirmed trend of better resolution on physics events at lower T. Set 11 mKas new working temperature; After towers installation, the cryostat was closed between September-November 2016: • cooldown started on Dec 5, 2016: lasted ~26 days (without counting technical stops for system debugging); • on Jan 26, 2017 reached a stable base temperature of T = 7 mK; cryogenics debugging electronics noise studies A. Branca

  12. Detector optimization • Developed tool to minimize vibrational noise from PTs: • use of linear drives (LD) to control PTs’ rotating valves; • PTs phase scan to find the configuration of minimum noise across the whole detector; • NTD resistances fixed by selecting the working temperature, while correct bias currents have been optimized with dedicated measurements: • for each bolometer set bias currents such to avoid distorted signal shapes and maximizing SNR; Reference pulses for different amplitudes @ 10 mK temperature: before and after LD switch on CUORE Preliminary A. Branca

  13. First data taking period and detector performance • First phase of detector optimization ended in April 2017 to start science runs: • Dataset 1: 3 weeks of data bracketed by 2 calibration periods (May - June 2017). TeO2 exposure ; • still room for performance improvements; • Second phase of detector optimization restarted in July 2017: • careful investigation/upgrades to the electronics grounding in the CUORE Faraday cage; • Introduced PTs phase scans to refine the abatement of induced noise; • optimization of the operating temperature and detector working points; • software and analysis upgrades; • Resumed data-taking in August 2017: • Dataset 2: same procedure as first dataset (August – September 2017). TeO2 exposure ; Dataset 1 Dataset 2 Improved resolution between the two datasets A. Branca

  14. Status of the CUORE experiment • Stop-A:warm-up to 100 K in order to fix a small leak (gate valves connecting DCS to cryostat) • Stop-B+C: • warm-up to 100 K; • solved small leaks and partial obstruction of condensation lines • fixed two PTs; Stop C Stop B Stop A • Introduced the Optimum-Trigger algorithm: allows to reduce the trigger thresholds; • Introduced an algorithm to reduce the correlated noise between bolometers; A. Branca

  15. From raw pulses to energy spectrum • Amplitude evaluation: Optimum Filter technique to maximize SNR; • Thermal gain stabilization: use reference pulses at constant energy (pulsers) to correct amplitude from small variations due to temperature drifts; • Calibration: apply calibration function, exploiting 6 known peaks from 232Th in calibration data. • Calibration data are also exploited for: • model detector response to energy deposition; • energy scale and resolution evaluation; • thermal gain stabilization; A. Branca

  16. Event selection • Noisy periods of data (about 1% of total exposure discarded); • Basic pulse quality selection (remove noise spikes, pile-up…); • Consider only events with multiplicity equal to 1 (anti-coincidence selection): • Radioactive contaminations likely produce signals in more then one crystal in a given time window; • Pulse shape analysis (PSA): • Define a distance in the 6-dimensional space of the pulse shape parameters; • Calculate a threshold maximizing the experimental sensitivity; • Discard events above the threshold; REMOVED ACCEPTED A. Branca

  17. fit results • A 3 components simultaneous UELM (Unbinned Extended Maximum Likelihood) fit in the energy region 2465-2575 keV is performed: • Posited peak at the 0v DBD Q-value: fixed peak position, floating rate common to all channels and both datasets; • Peak for 60Co sum gamma line (2505 keV): floating peak position, floating rate common to all channels and both dataset; • Flat background for multi scatter Compton events from 208Tl line and surface events: common to all channels and dataset dependent; • The peaks are modeled through the detector response function; • The line width for each peak is scaled by the resolution scaling factor obtained from calibration data; Signal decay rate best-fit: Background index best-fit (datasets average): A. Branca

  18. Combined limits with previous experiments Combined 90% limits on the half-life and the effective Majorana neutrino mass: CUORICINO: exposure ; CUORE-0: exposure ; CUORE: exposure • Half-life limits: • 130Te: 1.5 × 1025 yr from this analysis • 76Ge: 8.0 × 1025yr from PRL 120, 132503 (2018) • 136Xe: 1.1 × 1026 yr from Phys. Rev. Lett. 117, 082503 (2016) • 100Mo: 1.1 × 1024 yr from Phys. Rev. D 89, 111101 (2014) • 82Se: 2.4 × 1024 yr from Phys. Rev. Lett 120, 232502 (2018) • Nuclear Matrix elements from: • JHEP02 (2013) 025 • Nucl. Phys. A 818, 139 (2009) • Phys. Rev. C 87, 045501 (2013) • Phys. Rev. C 87, 064302 (2014) • Phys. Rev. C 91, 034304 (2015) • Phys. Rev. C 91, 024613 (2015) • Phys. Rev. C 91, 024309 (2015) • Phys. Rev. C 91, 024316 (2015) • Phys. Rev. Lett. 105, 252503 (2010) • Phys. Rev. Lett. 111, 142501 (2013) A. Branca

  19. Background: CUORE vs CUORE-0 • Excess of the 210Po peak: • under investigation; • not influencing the ROI counts ( < 1%); • In comparison to CUORE-0: • strong reduction in region: material selection/screening/cleaning; • consistent in almost all the region: same cleaning and assembly procedure; A. Branca

  20. Background reconstruction • Geant4 Monte-Carlo simulation: • Different background sources (60 components); • Surface/bulk contaminations of copper/lead/crystals; • Cosmic muons; • Background reconstructed with Bayesian fit on data exploiting JAGS: • Same data used for analysis; • Split in events crystal multiplicity: multiplicity 1 (M1), multiplicity 2 (M2) and multiplicity 2 sum (): • Each one is sensitive to different background contributions; • Geometrical split: outer/inner crystal layers: • Outer layer sensitive to cryostat contaminants; • Inner layer sensitive to tower contaminants; • Prior assigned to the normalization factor of each component: • From material screening, CUORE-0 background, • LNGS cosmic flux; M1 M2 inner outer A. Branca

  21. Background reconstruction Model able to reconstruct main features of the CUORE background A. Branca

  22. half-life Measured 130Te half life from CUORE: • Dominant systematic due to contamination location and uniformity; • Contribution from signal efficiency and energy thresholds are about one order of magnitude lower; Consistent with previous measurements: (CUORE-0) (NEMO-3) A. Branca

  23. half-life In CUORE events are of the the counts in the 1-2 MeV range because of reduced background In CUORE-0 events are of the counts in the 1-2 MeV range A. Branca

  24. Summary • With the first data (exposure of ): • set the strongest limit on the neutrino-less double-beta decay half-life of 130Te; • most precise measurement of the two neutrino double-beta decay half-life of 130Te; • Accumulated an exposure of more than to date: • exploring the possibility to lower the trigger thresholds; • physics results will be released shortly; • Gained a lot of experience from the operation of the first ton-scale bolometric array: • developed new methods for noise reduction; • refined methods to set the detector working conditions; • studying and testing new strategies to improve the detector resolution; • Future: • will take data for 5 years; • upgrades to perform particle identification (CUPID); A. Branca