1 / 40

Phakic IOL Overview

Phakic IOL Overview. António Marinho, MD PhD Departamento de Cirurgia Refractiva Hospital Arrábida Porto Portugal. DEFINITION. REFRACTIVE SURGERY To change in a permanent way the refractive power of the eye. How to achieve this goal ?. Change the corneal power (PRK,LASIK)

patb
Download Presentation

Phakic IOL Overview

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Phakic IOL Overview António Marinho, MD PhD Departamento de Cirurgia Refractiva Hospital Arrábida Porto Portugal

  2. DEFINITION • REFRACTIVE SURGERY • To change in a permanent way the refractive power of the eye

  3. How to achieve this goal ? • Change the corneal power (PRK,LASIK) • Change the power of the lens (RLE) • Introduce a new refractive surface (Phakic IOL)

  4. Why Phakic IOLs? Phakic IOL’s are ideal for high ametropias because: • High predictability even in very high ametropias • Stability of refraction • Preserve accomodation • No loss (usually gains) of lines of BSCVA

  5. Myopia - Subjective Refraction under - 7D : LASIK above -7D: Phakic IOL Main Factor : Pachymetry Hyperopia - Cycloplegic Refraction under + 3D : LASIK above + 4D: Phakic IOL Main factor: Keratometry Age Mínimal Age 18 years exceptions anisometropia Stable refraction in the last 18 months Above 50 years low ametropia LASIK high ametropia CLE WHEN PHAKIC IOLs ?

  6. INCLUSION CRITERIASpecific • Anterior chamber anatomy (AC depth and AC size) • Endothelium profile • Iris shape Pupil Size • Perfect Surgery

  7. AC Depth

  8. Bad Selection • Endothelial Decompensation • Shallow AC

  9. AC SIZE (OCT)

  10. Endothelium Profile • Endothelial cell count: • 21 to 25 years 2800 cells/mm • 26 to 30 years 2650 cells/mm • 31 to 35 years 2400 cells/mm • 36 to 45 years 2200 cells/mm • > 45 years 2000 cells/mm • Endothelial cell shape (avoid high polymagatism)

  11. Endothelial Cell Count • Before Surgery (inclusion criteria) • 3 months after (shows surgical trauma) • Yearly afterwards (if important decrease EXPLANT)

  12. ACRYSOF • Hydrophpbic Acrylic IOL • 4 point angle fixation • 6.0 mm Optic • -6.00/-16.50 • 4 sizes (12.5,13.0,13.5 and 14.0mm)

  13. Size Selection

  14. Acrysof Surgery • Introduce the IOL in the cartridge (diving position) • 2.6 mm incision • Inject the IOL into the eye (past pupil) • NO iridectomy • No suture

  15. ANGLE SUPPORTED AC PIOLs • Angle to angle distance very important • Size of the IOL is critical • Contact with the angle and iris root • May be close to endothelium • Far away from lens

  16. Rotation

  17. Rotation

  18. Peripheral synaechiae

  19. ARTISAN 5.0mm • Iris-Claw phakic IOL • PMMA • 5.0 mm O.Z. • Available for myopia, hyperopia (-23.00 to +12.00) and astigmatism( +/-)

  20. ARTISAN 6.0mm • Iris-Claw phakic IOL • PMMA • 6.0 mm O.Z. • Available for myopia (-2.00 to –15.00)

  21. Artisan Surgery • 2 side ports • Main incision • Fill AC with visco • Introduce and rotate the IOL • Enclavation of iris tissue • Iridectomy • Suture

  22. ARTIFLEX • Iris-claw phakic IOL • PMMA haptics • Silicone (foldable optic) • 6.00mm • One size fits all

  23. TORIC ARTIFLEX • Myopia -1.00 to -14.50 • Cylinder -1.00 to -7.50 • Two models (axis at 180º and 90º) • Sphere + Cylinder < -14.50

  24. ARTIFLEX • 2 side ports • Main incision (3.2mm) • Fill AC with visco • Introduce and rotate the IOL • Enclavation of iris tissue • Iridectomy • No Suture

  25. IRIS SUPPORTED PIOLs • One size fits all • No angle touch • Close contact with the iris (grasp) • Safe distance from the endothelium • Far away from the lens

  26. Not Perfect Surgery…. • Decentration is always a surgeon’s fault • These lenses are always centered regardless of the pupil • Luxation of the IOL(traumatic or spontaneous)is due to weak grasp

  27. Bad Selection • Posterior Synaechia • Convex Iris • Shallow AC

  28. IOL DEPOSITS • Rare • Disappear spontaneouly after 3 months in most cases • May need steroid treatment (exceptionally) • Related to surgical manipulation

  29. Posterior Chamber PIOLs

  30. ICL V4c • The NEW ICL V4 c has a tiny central hole in the middle of the optic • NO iridectomy is needed

  31. ICL Surgery • Load the ICL in the cartridge • 2 side ports (12 and 6) • Main incision (temporal) • Introduce IOL in AC • Place IOL behind the iris • Constricit the pupil • Iridectomy (if not YAG before)

  32. Posterior Chamber PIOLs • Sit on sulcus (ICL) or “float” in aquous humour (PRL) • “Vault” (the space between ICL and lens) is crucial and depends on the IOL size • Close contact with the lens • Very far away from the endothelium

  33. Size matters….. • Short ICL: Decentration and small vault • Long IOL: Excessive vault

  34. If there is no vault… • Anterior subcapsular cataract (less frequent as the surgical technique and sizing devices get better)

  35. Refractive ResultsBCVA>20/40 • Artisan 93.9% (518 eyes) • ICL 94.7% (331 eyes) • Cachet 100% (113 eyes)

  36. Refractive Results Safety • PIOL GAIN LOSS • Artisan 43.5 % 1.2% • ICL 40.6% 0% • Cachet 27.3% 0%

  37. AVAILABILITY

  38. Inclusion criteriaPIOLs

  39. PIOLs Surgery Overview

  40. Refractive ResultsConclusions • All Phakic IOLs have GREAT refractive results • Most eyes gain lines • The KEY to select a phakic IOL are not the refractive results ,but the complications