1 / 9

Community Preparation for Caring for Mechanical Circulatory Device Patients

Community Preparation for Caring for Mechanical Circulatory Device Patients. University of Wisconsin Hospital And Clinics Ventricular Assist Device Program. Mechanical Circulatory Support Device Overview. MCDSs are devices to support the failing heart (one side or both)

qiana
Download Presentation

Community Preparation for Caring for Mechanical Circulatory Device Patients

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Community Preparation for Caring for Mechanical Circulatory Device Patients University of Wisconsin Hospital And Clinics Ventricular Assist Device Program

  2. Mechanical Circulatory Support Device Overview • MCDSs are devices to support the failing heart (one side or both) • Most MCDS patients are anticoagulated with Coumadin and Aspirin • All MCDSs have: • Pump (implanted inside the patient’s chest) • Computer to control the settings/ display VAD performance readings • Power source: batteries or console connected to AC power

  3. Heartware HVAD Components Small pump attaches directly to heart Thin, flexible driveline cable exits skin A small controller & batteries run the pump

  4. Heartware HVAD Readings 3 Readings: Flow= cardiac output from the pump Target varies by patient Speed= how fast pump is running rpm - range 2400-3200 rpm Power= how much watts it takes to run the pump - normal is < 8 watts

  5. Heartware HVAD Care Overview • Cannot palpate a pulse • Pulse oximeter may/ may not work • Need to assess if pump is running: listen over the point of maximal impulse (PMI) area for a consistent mechanical hum • Blood pressure: need a Doppler and sphygmomanometer • result is considered a MAP • acceptable MAP 60-85 mmHg

  6. Heartware HVAD Care Overview • EKG conduction-looks normal • does not impact VAD function unless RV fails • VAD will continue to receive blood supply until RV fails or pulmonary hypertension • Most of these patients have an ICD/ pacer. Some patients’ ICDs are off to avoid inappropriate shocks. • Most VAD patients tolerate the arrhythmias well and may have little symptoms for period of time. • Assess for patient’s tolerance of situation (if able). Ask for symptoms: lightheadedness, shortness of breath, palpitations, bloating, pain (anywhere), fever, chills, diaphoresis

  7. Heartware HVAD Controller Faceplate

  8. EMERGENCY CARE • Call implant hospital: 608-263-6400 and ask for HEART FAILURE ATTENDING ON CALL • Still need to follow Airway Breathing of ABC’s • √ VAD stopped (listen over PMI area for consistent mechanical hum). If stopped significant risk for clot inside MCDS and thromboembolize. • No chest compressions • Defibrillation- STANDARD METHOD • Heparin bolus if VAD stopped (contact implant hospital first)

  9. Contacts 608-263-6400 (Paging Operator) • Heart Failure Attending • Ventricular Assist Device Coordinator (non-emergency)

More Related