An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. FRCR Physics Lectures The Physics of Diagnostic Ultrasound Mark Wilson Radiotherapy Physicist, Queen’s Centre for Oncology

  2. Session 3 Overview Session Aims: Recap Image Artefacts Contrast Agents Introduction to Doppler Ultrasound

  3. Quick Recap

  4. Ultrasound Waves  = c / f The term Ultrasound refers to high frequency sound waves. Sounds waves are mechanical pressure waves which propagate through a medium causing the particles of the medium to oscillate backward and forward The velocity and attenuation of the ultrasound wave is strongly dependent on the properties of the medium through which it is travelling c =  k / 

  5. Pulse-Echo Principal and Image Acquisition D Source of sound ) Sound reflected at boundary ) ) ) Pulse ) Distance = Speed x Time 2D = c x t Echo Diagnostic ultrasound utilises the pulse-echo principle Each pulse-echo sequence produces one line of the image Several pulse-echo sequences are needed to compose a full image frame.

  6. Ultrasound Interactions with Matter Ultrasound waves undergo the following interactions: Reflection Scatter Refraction Attenuation and Absorption

  7. ( ) 2 Z2 – Z1 Ir R = = Ii Z1 + Z2 Ii Reflection Z1 Ir Intensity Reflection Coefficient (R) Z2 It Acoustic Impedance z =  k Acoustic Impedance z = c

  8. Strength of reflection depends on the difference between the Z values of the two materials • Ultrasound only possible when wave propagates through materials with similar acoustic impedances – only a small amount reflected and the rest transmitted • Therefore, ultrasound not possible where air or bone interfaces are present Reflection

  9. Scattering Reflection occurs at large interfaces such as those between organs where there is a change in acoustic impedance Within most organs there are many small scale variations in acoustic properties which constitute small scale reflecting targets Reflection from such small targets does not follow the laws of reflection for large interfaces and is termed scattering Scattering redirects energy in all directions, but is a weak interaction compared to reflection at large interfaces

  10. Reflection (echo) Incident Refraction θi θr Z1 =1c1 sin θt c2 When an ultrasound wave crosses a tissue boundary at an angle (non-normal incidence), where there is a change in the speed of sound c, the path of the wave is deflected as it crosses the boundary Z2 =2c2 = c1 sin θi θt Transmission (refraction)

  11. 1 Attenuation Relative Intensity, I Low frequency 0.5 High frequency As an ultrasound wave propagates through a medium, the intensity reduces with distance travelled Attenuation describes the reduction in intensity with distance and includes scattering, diffraction, and absorption Attenuation increases linearly with frequency Limits frequency used – trade off between penetration depth and resolution Distance travelled, d I = Ioe- d Where  is the attenuation coefficient

  12. Attenuation Absorption In soft tissue most energy loss (attenuation) is due to absorption Absorption is the process by which ultrasound energy is converted to heat in the medium Absorption is responsible for tissue heating DecibelNotation Attenuation and absorption is often expressed in terms of decibels Decibel, dB = 10 log10 (I2 / I1)

  13. Image Artefacts

  14. Artefacts When forming a B-mode image, a number of assumptions are made about ultrasound propagation in tissue. These include: Speed of sound is constant Attenuation in tissue is constant Ultrasound pulse travels only to targets that are on the beam axis and back to the transducer Significant variations from these conditions in the target tissues are likely to give rise to visible image artefacts

  15. Range Errors The distance, d, to the target is derived from the time elapsed between transmission of the pulse and receipt of the echo from the target, t In making this calculation the system assumes that t = 2d / c, where the speed of sound is constant at 1540 m/s If the speed of sound in the medium between the transducer and target is greater (or less) than 1540 m/s, the echo will arrive back at the transducer earlier (or later) than expected for a target of that range Tissue c = 1540 m/s Fat c = 1420 m/s Target Displayed at

  16. Target Displayed at Refraction Medium 1 c1 Refraction of the ultrasound beam as it passes between tissues with a different speed of sound can result in objects appearing at an incorrect position in the image Medium 2 c2 > c1

  17. Attenuation Artefacts During imaging the outgoing pulse and returning echoes are attenuated as they propagate through tissue, so that echoes from deeper targets are weaker than those from similar superficial targets Time Gain Compensation (TGC) is applied to correct for such changes in echo amplitude with target depth Most systems apply a constant rate of compensation designed to correct for attenuation in typical uniform tissue The operator can also make additional adjustments to compensate via slide controls that adjust the gain applied specific depths in the image TGC artefacts may appear in the image when the applied compensation does not match that actual attenuation rate in the target tissue

  18. Image of renal cyst Acoustic Enhancement Occurs when ultrasound passes through a tissue with low attenuation Echoes from deeper lying tissues are enhanced due to the relatively low attenuation in the overlying tissue This occurs because the TGC is set to compensate for the greater attenuation in the adjacent tissues Low attenuation

  19. Image of Gallstone Acoustic Shadowing High attenuation Occurs when ultrasound wave encounter a very echo dense (highly attenuating) structure Nearly all of the sound is reflected, resulting in an acoustic shadow This occurs because the TGC is set to compensate for the lower attenuation in the adjacent tissues

  20. Transducer Reverberation Artefact Interface Reverberation artefacts arise due to reflections of pulses and echoes by strongly reflecting interfaces Occur most commonly where there is a strongly reflecting interface parallel to the transducer face Involves multiple reflections - Initial echo returns to reflecting interface as if it is a weak transmission pulse and returns a second echo (reverberation) Reverberation

  21. Reverberation Artefacts

  22. Reverberation Artefacts

  23. Contrast Agents

  24. Ultrasound ContrastAgents Ultrasound contrast agents are gas-filled micro-bubbles which are injected into the blood stream Micro-bubbles will give increased backscatter signal due to the large acoustic impedance mismatch between the gas-filled bubble and surrounding tissue

  25. Ultrasound ContrastAgents Micro-bubble suspension is injected intravenously into the systemic circulation in a small bolus The micro-bubbles will remain in the systemic circulation for a certain period of time Ultrasound waves are directed on the area of interest and when the micro-bubbles in the blood flow past the imaging window they give rise to increased signal Allows detection of blood flow where it would otherwise not be seen

  26. Ultrasound ContrastAgents

  27. Targeted ContrastAgents Targeted contrast agents are under preclinical development They retain the same general features as untargeted micro-bubbles, but they are outfitted with ligands that bind to specific receptors expressed by cell types of interest Micro-bubbles theoretically travel through the circulatory system, eventually finding their respective targets and binding specifically If a sufficient number of micro-bubbles have bound to the target area, an increased signal will be seen

  28. Doppler Ultrasound

  29. The Doppler Effect The Doppler effect is observed regularly in our daily lives, e.g. it can be heard as the changing pitch of an ambulance siren as it passes by The Doppler effect is the change in the observed frequency of the sound wave (fr) compared to the emitted frequency (ft) which occurs due to the relative motion between the observer and the source Consider three situations - Source and observer stationary - Source moving towards observer - Source moving away from observer

  30. ) ) Observer ) ) ) The Doppler Effect fr = ft Source The observed sound has the same frequency as the emitted sound Source and observer stationary (Note: Frequency is the number of cycles per second)

  31. ) ) ) ) ) The DopplerEffect fr > ft Causes the wavefronts travelling towards the observer to be more closely packed, so that the observer witnesses a higher frequency wave than emitted Source moving towards observer

  32. ) ) ) ) ) The DopplerEffect fr < ft The wavefronts travelling towards the observer will be more spread out, so that the observer witnesses a lower frequency wave than emitted Source moving away from observer

  33. The Doppler Effect The resulting change in the observed frequency from that transmitted is known as the Doppler shift The magnitude of the Doppler shift frequency is proportional to the relative velocity between the source and the observer It does not matter if it is the source or the observer is moving The Doppler effect enables Ultrasound to be used to assess blood flow by measuring the change in frequency of the ultrasound scattered from moving blood

  34. Ultrasound Measurement of Blood Flow Transducer is held stationary and the blood moves with respect to the transducer The ultrasound waves transmitted by the transducer strike the moving blood, so the frequency of the ultrasound experienced by the blood is dependent on whether the blood is stationary, moving towards or away from the transducer The blood then scatters the ultrasound, some of which travels in the direction of the transducer and is detected The scattered ultrasound is Doppler frequency shifted again as a result of the motion of the blood, which now acts as a moving source Therefore, a Doppler shift has occurred twice between the ultrasound being transmitted and received back at the transducer

  35. cos 2 ft v Doppler frequency shift, fd = fr – ft = Ultrasound Measurement of Blood Flow c Target direction  fr = received frequency ft = transmitted frequency c = speed of sound v = velocity of blood  = angle between the path of the ultrasound beam and the direction of the blood flow (angle of insonation)

  36. cos 1 Ultrasound Measurement of Blood Flow The detected Doppler shift also depends on the cosine of the angle  between the path of the ultrasound beam and the direction of blood flow The operator can alter  by adjusting the orientation of the transducer on the skin surface Desirable to adjust  to obtain the highest Doppler frequency shift  0 90

  37. Ultrasound Measurement of Blood Flow c fd v = 2 ft cos If the angle of insonation of the ultrasound beam is known it is possible to use the Doppler shift frequency to estimate the velocity of the blood using the Doppler equation In diseased arteries the lumen will narrow and the blood velocity will increase

  38. Narrowing in artery Ultrasound Measurement of Blood Flow V1 V2 A2 A1 The flow (Q) remains constant Q =A1 V1 = A2 V2 A = Area V = Velocity

  39. Thank youAny Questions?

  40. Session4Overview Session Aims: Continuous Wave Doppler US Pulsed Wave Doppler US Harmonic Imaging

  41. Doppler Ultrasound Continued

  42. Continuous Wave and Pulsed Wave Doppler Doppler systems can be either continuous wave or pulsed wave: Continuous wave (CW) systems transmit ultrasound continuously Pulsed wave (PW) systems transmit short pulses of ultrasound The main advantage of PW Doppler is that Doppler signals can be acquired from a known depth The main disadvantage of PW Doppler is that there is an upper limit to the Doppler frequency shift which can be detected

  43. Continuous Wave (CW) Doppler In a CW Doppler system there must be separate transmission and reception of ultrasound – transducer with two separate elements The region from which Doppler signals are obtained is determined by the overlap of the transmit and receive ultrasound beams

  44. Pulsed Wave (PW) Doppler Transducer Gate depth Gate length In a PW Doppler system it is possible to use the same transducer element for both transmit and receive The region from which Doppler signals are obtained is determined by the depth of the gate and the length of the gate, which can both be controlled by the operator

  45. Ultrasound signal received by transducer The received ultrasound signal consists of the following four types of signal: echoes from stationary tissue echoes from moving tissue echoes from stationary blood echoes from moving blood The task for the Doppler system is to isolate and display the Doppler signals from blood, and remove those from stationary and moving tissue

  46. Amplitude Ultrasound signal received by transducer Tissue Doppler signals from blood tend to be low amplitude (small reflected echo) and high frequency shift (high velocity) Doppler signals from tissue are high amplitude (large reflected echo) and low frequency shift (low velocity) These differences provide the means by which signals from true blood flow may be separated from those produced by surrounding tissue Blood Frequency

  47. Demodulation Separation of the Doppler frequencies from the underlying transmitted signal Transducer Doppler signal processing Demodulator High-pass filtering Removal of the tissue signal Signal processor High-pass filter Frequency estimation Calculation of Doppler frequency and amplitudes Frequency estimator Display

  48. Demodulation The Doppler frequencies produced by moving blood are a tiny fraction of the transmitted ultrasound frequency E.g. If transmitted frequency is 4 MHz, a motion of 1 m/s will produce a Doppler shift of 5.2 kHz, which is less the 0.1% of the transmitted frequency The extraction of the Doppler frequency information from the ultrasound signal received from tissue and blood is called demodulation In PW Doppler, need the PRF to be at least twice the maximum Doppler shift frequency in order to avoid ‘aliasing’ (not a problem in CW) Aliasing is an artefact introduced by under-sampling in which high frequency components take the alias of a low frequency component

  49. Amplitude Amplitude High-pass Filtering High-pass Filtering Tissue Blood Blood Frequency Frequency

  50. Frequency Estimation Frequency shift A spectrum analyser calculates the amplitude of all the frequencies present within the Doppler signal In the spectral display the brightness is related to the amplitude of the Doppler signal component at that particular frequency Time