An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Field Evaluation of Diffusive Samplers for Indoor Air VOC Measurements AIHce 2011 Heidi Hayes, Technical Director Robert Mitzel, Vice-President Business Development 1201

  2. Outline • Introduction • Study Objectives • Sampler Selection • Field Sampling • Results • Conclusions

  3. Conventional Air Sampling • Possible equipment failure • Requires experienced field sampler • Short duration (~24 hours) • Expensive to ship • Requires experienced field sampler • Short duration (~8 hours) • Sorbent type and sample volume selection is critical Summa Canisters Pumped Sorbent Tubes

  4. Passive Sampling Practical Advantages • Reliable deployment with little training required • Unobtrusive • Inexpensive to ship Technical Advantages • Capable of generating trace level RLs • Long-term time-integrated measurements More representative indoor air concentrations and increased sensitivity are advantageous to health risk assessments.

  5. Passive Sampling Concepts Measured in lab Analytical Result (µg) 1000 mL 1L 1000 L m3 X X Concentration (µg/m3) Uptake Rate (mL/min) Sampling duration (min) X Available in literature Dependent on Sampler Geometry Recorded in the field

  6. Project Objectives • Sample integration of 1 to 7 days • Measurement of a wide VOC suite • Petroleum and chlorinated compounds • Reporting limits comparable to TO-15 SIM (~0.1 µg/m3) • Measured concentrations correlate with TO-15

  7. Passive Samplers Sampler Geometries Tube Membrane Badge Radial Analytical Sensitivity Sorbent Sampling Rate Solvent WMS®- Charcoal SKC 575 3M OVM Radiello®130 Thermal Desorption ATD WMS®- TD SKC ULTRA Radiello®145 Lowest Highest

  8. Passive Samplers Sampler Geometries Tube Membrane Badge Radial Sorbent Solvent Radiello®130 Thermal Desorption SKC ULTRA Radiello®145

  9. Field Sampling – Case 1 • Indoor air samples collected • Duration 3, 4, and/or 7 days • Concurrent deployment • Radiello 130 – Charcoal • Radiello 145 – TD Sorbent • ULTRA III – TD sorbent

  10. Results Good comparability was observed when detections on each sampler were sufficiently above their respective reporting limits.

  11. Results • ULTRA III = 5-20 times greater sensitivity than the RAD-Charcoal. • ULTRA III had validated sampling rates for chlorinated breakdown products whereas RAD-TD required estimated rates. • Diffusive adsorption on the RAD-TD sorbent did not behave as predicted for these light VOCs (chloroform, 1,1-DCE) resulting in low bias. Stronger TD sorbent is required for these VOCs.

  12. Results • One indoor air location was severely impacted with chlorinated solvents (100 to 10,000 µg/m3) • Sampling duration was 3 days. • Both Radiello-TD and ULTRA III exceeded capacity & TD-GC/MS. • Radiello-Charcoal had a higher capacity, and solvent extraction allowed for easy dilutions.

  13. Field Sampling – Case 2 • Indoor air samples collected • 13 sites • Concurrent TO-15 cans & ULTRA III • Chlorinated solvents,petroleum products • 1 to 3 day duration

  14. Results Strong correlation between ULTRA III and TO-15 concentrations across 3 orders of magnitude and at concentrations <0.1 µg/m3

  15. Conclusions • Each passive sampler evaluated provided quantitative VOC indoor air measurements for TCE and PCE over a period of up to 7 days. • The larger surface area of charcoal provided an advantage over TD-sorbents when sampling high concentrations over multiple days.

  16. Conclusions • ULTRA III-TD and Radiello-TD provide greater sensitivity than the Radiello-Charcoal over the 1-7 day period. • ULTRA III-TD provides a wider range of VOCs than Radiello-TD. • ULTRA III has a built-in blank correction allowing for improved accuracy at trace levels.